Document Type : Original Article(s)

Authors

1 Cardiologist, Fellowship of Echocardiography, Associate Professor, Department of Cardiology, Najafabad Branch, Islamic Azad University, Najafabad, Iran

2 General Practitioner, Najafabad Branch, Islamic Azad University, Najafabad, Iran

Abstract

BACKGROUND: Accuracy of blood pressure (BP) measurement in clinical settings is one of the most concerns despite of promotion in techniques for the measurement of BP. Our aim was to compare automated versus manual BP measurement in intensive care unit (ICU), coronary care unit (CCU), and emergency room patients. METHODS: Totally, 117 patients in ICU, CCU, and emergency department were registered in the study. Demographic information was recorded. The cardioset heart monitoring device was used for measuring BP and mercury sphygmomanometer with appropriate cuffs was used for manual method. Then, the mean BP of two methods was compared based on different age, sex, weight, and disease findings. RESULTS: The mean systolic blood pressure (SBP) was 124.526 mmHg, with minimum and maximum of 123.111 and 125.940 mmHg, respectively (Cronbach’s alpha = 0.893); furthermore, mean diastolic blood pressure (DBP) was 73.496 mmHg, with minimum and maximum of 72.718 and 74.247 mmHg, respectively (Cronbach’s alpha = 0.852). SBP was significantly different between the two methods, and especially in patients below 60 years, hospitalized in ICU ward, overweight, mid-upper arm circumference below 27 cm, and with neurosurgery problems, it was higher by manual method, too (P < 0.050). Moreover, DBP was more in a manual method in patients with female sex, below 60 years, hospitalized in ICU ward and with neurosurgery problems (P < 0.050). CONCLUSION: The results of this study suggested that manual method in measurement of BP frequently shows higher BP, especially in patients admitted to hospitals-affecting up to 15 mmHg higher, and this discrepancy is more in critical situations. 

Keywords

  1. Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, et al. Recommendations for blood pressure measurement in humans and experimental animals: part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Circulation 2005; 111(5): 697-716.
  2. Myers MG, Godwin M, Dawes M, Kiss A, Tobe SW, Kaczorowski J. Measurement of blood pressure in the office: Recognizing the problem and proposing the solution. Hypertension 2010; 55(2): 195-200.
  3. Sala C, Santin E, Rescaldani M, Magrini F. How long shall the patient rest before clinic blood pressure measurement? Am J Hypertens 2006; 19(7): 713-7.
  4. Giles TD, Egan P. Pay (adequately) for what works: The economic undervaluation of office and ambulatory blood pressure recordings. J Clin Hypertens (Greenwich) 2008; 10(4): 257-9.
  5. Myers MG, Oh PI, Reeves RA, Joyner CD. Prevalence of white coat effect in treated hypertensive patients in the community. Am J Hypertens 1995; 8(6): 591-7.
  6. Beckett L, Godwin M. The BpTRU automatic blood pressure monitor compared to 24 hour ambulatory blood pressure monitoring in the assessment of blood pressure in patients with hypertension. BMC Cardiovasc Disord 2005; 5(1): 18.
  7. Myers MG, Valdivieso M, Kiss A. Use of automated office blood pressure measurement to reduce the white coat response. J Hypertens 2009; 27(2): 280-6.
  8. Parati G, Omboni S, Bilo G. Why is out-of-office blood pressure measurement needed? Hypertension 2009; 54(2): 181-7.
  9. Verdecchia P, Angeli F, Mazzotta G, Gentile G, Reboldi G. home blood pressure measurements will not replace 24-hour ambulatory blood pressure monitoring. Hypertension 2009; 54(2): 188-95.
  10. Staessen JA, Byttebier G, Buntinx F, Celis H, O'Brien ET, Fagard R. Antihypertensive treatment based on conventional or ambulatory blood pressure measurement. A randomized controlled trial. Ambulatory Blood Pressure Monitoring and Treatment of Hypertension Investigators. JAMA 1997; 278(13): 1065-72.
  11. Wright JM, Mattu GS, Perry Jr TL, Gelferc ME, Strange KD, Zorn A, et al. Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor. Blood Press Monit 2001; 6(3): 161-5.
  12. White WB, Anwar YA. Evaluation of the overall efficacy of the Omron office digital blood pressure HEM-907 monitor in adults. Blood Press Monit 2001; 6(2): 107-10.
  13. Stergiou GS, Tzamouranis D, Protogerou A, Nasothimiou E, Kapralos C. Validation of the Microlife Watch BP Office professional device for office blood pressure measurement according to the International protocol. Blood Press Monit 2008; 13(5): 299-303.
  14. Suokhrie LN, Reed CR, Emory C, White R, Moriarity CT, Mayberry J. Differences in automated and manual blood pressure measurement in hospitalized psychiatric patients. J Psychosoc Nurs Ment Health Serv 2013; 51(3): 32-7.
  15. Myers MG, Godwin M, Dawes M, Kiss A, Tobe SW, Kaczorowski J. The conventional versus automated measurement of blood pressure in the office (CAMBO) trial: Masked hypertension sub-study. J Hypertens 2012; 30(10): 1937-41.
  16. Myers MG, Godwin M, Dawes M, Kiss A, Tobe SW, Kaczorowski J. Conventional versus automated measurement of blood pressure in the office (CAMBO) trial. Fam Pract 2012; 29(4): 376-82.
  17. Myers MG, Godwin M, Dawes M, Kiss A, Tobe SW, Grant FC, et al. Conventional versus automated measurement of blood pressure in primary care patients with systolic hypertension: Randomised parallel design controlled trial. BMJ 2011; 342: d286.
  18. Heinemann M, Sellick K, Rickard C, Reynolds P, McGrail M. Automated versus manual blood pressure measurement: A randomized crossover trial. Int J Nurs Pract 2008; 14(4): 296-302.
  19. O'Brien E. Has conventional sphygmomanometry ended with the banning of mercury? Blood Press Monit 2002; 7(1): 37-40.
  20. Scientific Committee on Emerging and Newly Identified Health Risks SCENIHR. Mercury sphygmomanometers in healthcare and the feasibility of alternatives [Online]. [cited 2009]; Available from: URL: http://ec.europa.eu/health/ph_risk/ committees/04_scenihr/docs/scenihr_o_025.pdf
  21. Myers MG. Automated blood pressure measurement in routine clinical practice. Blood Press Monit 2006; 11(2): 59-62.
  22. Head GA, Mihailidou AS, Duggan KA, Beilin LJ, Berry N, Brown MA, et al. Definition of ambulatory blood pressure targets for diagnosis and treatment of hypertension in relation to clinic blood pressure: Prospective cohort study. BMJ 2010; 340: c1104.
  23. Myers MG, Valdivieso MA. Use of an automated blood pressure recording device, the BpTRU, to reduce the "white coat effect" in routine practice. Am J Hypertens 2003; 16(6): 494-7.
  24. Myers MG, Valdivieso M, Kiss A. Consistent relationship between automated office blood pressure recorded in different settings. Blood Press Monit 2009; 14(3): 108-11.